FTIR QUANTITATIVE ANALYSIS OF LOW CONCENTRATION FURALDEHYDE
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    This paper studies the determination methods offuraldehyde and compares them with each other in a furaldehyde-benzenesample system that contains less than 10% of furaldehyde.The recentChinese national standard method is that of hydroxylamine(GB 1926-80).This method has an error approximately greater than 5%,and it has moresteps and is time-consuming.The Nicolet DX quantitative analysis packagecauses even greater error,while FTIR method has errors less than 5%by selecting proper baselines and drawing a work curve.FTIR methodis simple,quick and nondestructive,and needs only few amounts ofsamples.It's suitable to the samples of the system of the experiment.

    Reference
    Related
    Cited by
Get Citation

郑泽根,张玉奇.低含量糠醛的红外光谱定量分析[J].土木与环境工程学报(中英文),1991,13(2):