SEVERAL THEORETICAL PROBLEMS OF VISCOELASTICITY MEASUREMENT OF BLOOD
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Q66

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    In order to tackle the present difficult problems in viscoelasticity measurement of blood and other biofluids with Couette and cone-plate rheometers, the effect of inertia of inner cylinder and sample on the measurement result is studied. Rheological models of the instruments are reconsidered. Based on theoretical analysis of the flow state of the non-Newtonian fluid,a corrected method for calculating Theological parameters of sample is proposed.

    Reference
    Related
    Cited by
Get Citation

蔡绍皙 吴西.水的弹性分量困惑和血液粘弹性测量[J].重庆大学学报,1991,14(4):73~77