Value traceability methods and metrological calibration devices for digital image correlation techniques
Author:
Affiliation:

1.Chongqing Academy of Metrology and Quality Inspection, Chongqing 401123, P. R. China;2.Chongqing Airport Group Co., Ltd., Chongqing 401120, P. R. China

Fund Project:

Supported by Chongqing Research Institution Performance Incentive Guidance Special Project(cstc2020jxj1X0004).

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    针对当前数字图像相关法存在校准方法缺乏系统性、量值溯源体系不完善的问题,提出一种基于激光干涉测量技术的校准方法,并结合自主设计局部变形测量光路系统实现精准测量。通过有限元数值模拟,系统研究了拉伸荷载下应变板的应力刚化效应。基于理论分析与试验验证,建立了环境温度荷载作用下的应变板热-力耦合变形模型,量化了应变板表面对流换热性能对应变测量结果的影响。通过引入阿贝误差补偿机制,有效抑制了应变板拉伸过程中极微小弯曲变形导致的非线性误差。此外,系统分析了测量过程中各影响量引入的标准不确定度分量及校准装置整体不确定度,并完成了应变校准试验。结果表明,所提出的基于激光干涉原理的数字图像相关法校准装置可将标准应变场的测量不确定度降低至1.9%~0.83%。

    Abstract:

    The digital image correlation (DIC) method currently suffers from inadequate systematic calibration methods and an underdeveloped metrological traceability system. To address these challenges, this study proposes a novel calibration methodology based on laser interferometry, integrated with a custom-designed optical system for local deformation measurements. Finite element simulations are employed to analyze the stress stiffening effect in strain plates under tensile loading. A coupled thermo-mechanical model is developed via theoretical analysis and experimental validation to quantitatively assess the influence of convective heat transfer on surface strain measurement accuracy. Furthermore, an Abbe error compensation mechanism is implemented to effectively mitigate nonlinear measurement errors induced by microscale bending during loading. A comprehensive uncertainty analysis is conducted to evaluate both standard uncertainty components and the overall measurement uncertainty. Experimental results demonstrate that the proposed calibration device reduces the uncertainty in standard strain field measurements to the range of 1.9% to 0.83%, offering a critical advancement toward establishing a traceable calibration framework for DIC-based strain measurement systems.